Information was received from a manufacturer representative regarding a patient who was implanted with a neurostimulator.It was reported that the patient attempted to turn on the implantable neurostimulator (ins) last week and could not.The manufacturer representative interrogated the ins on (b)(6) 2016 with the clinician programmer and got the message that a reset occurred and the serial number was gone.The session was able to be continued and the battery information was run.The voltage was 2.52v and capacity used was 45-90%.Impedances at both 1.5v and 3v were run and ¿???¿ were seen at the combinations 0/4, 0/5, 1/5, 2/3, and 3/5.The patient reported no falls or traumas but they had started low impact aerobics and yoga in (b)(6).The patient had recent exposure to security gates and theft detectors in the airport security sometime in (b)(6).The manufacturer representative did threshold testing with the patient, activating most of the electrodes.However, when this was done the voltage was at 10.5v and the patient was not feeling anything except a zinging in their knee.The patient felt a zinging sensation in their ankle during impedance testing.The last time the patient felt stimulation normally was in (b)(6).Impedances were checked several more times with the ¿???¿ continuing to show up on several combinations.The measurement for 4-7 was showing 70 ohms.Prior to one of the impedance tests a power on reset was seen, prompting the manufacturer representative to enter the serial number.The manufacturer representative entered the serial number into the clinician programmer and ran impedances again with 4v obtaining the following results: 0-1 at ???, 0-2 at 529, 0-3 at ???, 0-4 at ???, 0-5 at 2084, 0-6 at 671, 0-7 at 735, 1-2 at 1176, 1-3 at 1171, 1-4 at 1661, 1-5 at 2787, 1-6 at 1661, 1-7 at 1661, 2-3 at 529, 2-4 at 907, 2-5 at 2084, 2-6 at 813, 2-7 at 907, 3-4 at 671, 3-5 at 2084, 3-6 at 671, 3-7 at 2084, 4-5 at 1661, 4-6 at 813, 4-7 at 104, 5-6 at ???, 5-7 at 1661, and 6-7 at 735.The manufacturer representative tested 2-3 combinations with 210 pulse width and 30 rate.At 4.4v the patient started to feel it but it was ¿static-y¿ and not normal like it had been.The patient was previously programmed on electrodes 0-3 and 4-7 for programs 1 and 2.It was reported that the patient was reprogrammed using the contacts that were within range.The patient had discussed lead revision/replacement with the health care professional (hcp).
|