| Model Number |
P10.7-80D335 |
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The patient was implanted on (b)(6) 2016.The nail was removed on (b)(6) 2016 and the patient was implanted with a new nail without incident.To date, the patient is doing fine and no negative outcomes were reported.A preliminary evaluation from x-rays provided revealed that the internal components of the nail appear to have no damage to prevent the nail from distracting; the nail appeared to have lengthened.A dhr review revealed that the device met all of the required quality inspections, and was released within specifications.
|
| |
|
Event or Problem Description
|
|
A distributor reported that a patient's precice nail was removed approximately three (3) weeks of implantation; the nail allegedly appeared to not be lengthening.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection of the returned device revealed that the nail was undamaged.X-ray images revealed no damage to the internal components.Functional testing of the precice nail revealed that the device was able to be distracted and retracted when tested with both a high speed magnet tool and an erc unit.The device operated within specifications and was fully functional; no malfunction was detected.A review of the lot history record revealed that the nail met all the required quality inspections and was released within specifications.
|
| |
|