A.I.D.D LONGFORD ARCHITECT TSH Back to Search Results
Catalog Number 07K62-35
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/19/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed a falsely depressed thyroid stimulating hormone (tsh) result for one patient on the architect i2000sr analyzer.The following data was provided sid (b)(4): initial between 60-80, repeat (falsely depressed sid (b)(4)) 17.86, repeated the next day sid (b)(4)= 63.There is no known impact to patient management.
 
Additional Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing.Historical performance of the reagent lot was evaluated using world wide data from abbott link.The patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for this lot is within the established control limits and no unusual reagent lot performance was identified.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The product was not available for return.An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications.Based on all available information and abbott diagnostics' complaint investigation, the assay performed as intended and no product deficiency was identified.
 
Additional Manufacturer Narrative
The conclusion code was corrected.The device evaluation was reassessed and concluded that a malfunction occurred, therefore, the device was not performing as intended, however, no systemic issue or product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TSH
Common Device Name
TSH
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford NA
EI  NA
MDR Report Key5930810
Report Number3005094123-2016-00045
Device Sequence Number1320070
Product Code JLW
UDI-Device Identifier00380740105969
UDI-Public00380740105969
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2017
Device Catalogue Number07K62-35
Device Lot Number64210UI00
Was Device Available for Evaluation? Yes
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 08/19/2016
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date09/06/2016
Supplement Report FDA Received Date09/13/2016
05/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT I2000SR ANALYZER; LN 03M74-02, SN (B)(4)
-
-