ENCORE MEDICAL, L.P. 3DKNEE; INSERT, 3D EX SZ 10RT 9MM Back to Search Results
Catalog Number 392-09-710
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/10/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery due to an infection.There was an antibiotic spacer in the patient, all of the components had been removed previously.
 
Additional Manufacturer Narrative
The reason for this revision surgery was an infection.The previous surgery and the revision detailed in this investigation occurred over 3 years apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.The healthcare professional indicated there was a serious risk to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) and product complaint report (pcr) database records show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DKNEE
Common Device Name
INSERT, 3D EX SZ 10RT 9MM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key5930854
Report Number1644408-2016-00611
Device Sequence Number814604
Product Code OIY
UDI-Device Identifier00888912120296
UDI-Public(01)00888912120296
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K091956
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/30/2017
Device Catalogue Number392-09-710
Device Lot Number59602400
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 08/11/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/06/2016
Supplement Report FDA Received Date10/06/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
130-03-035, LOT 476A1000; 130-03-035, LOT 476A1045; 200-05-008, LOT 39491L01; 200-10-008, LOT 699A1006; 206-01-008, LOT 758A1006; 210-05-008, LOT 768A1017; 210-05-008, LOT 774A1005; 220-15-022, LOT 872A1003; 320-15-016, LOT 159B1003; 322-01-108, LOT 53897487; 360-11-508, LOT 59601571
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age54 YR
Patient Weight102
-
-