BIOSENSE WEBSTER, INC. (JUAREZ); CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number D-1260-04-S
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Cardiac Tamponade (2226)
Date of Event 08/09/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the product was not returned for analysis, no product failure analysis can be conducted and no determination of possible contributing factors could be made.As such, the investigation will be closed.If the complaint device is received in the future we will reopen the complaint and perform the investigation as appropriate.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Concomitant products: carto 3 system, model # m-4800-01, s/n: (b)(4).Smartablate generator (1 of 2), model # m-4900-07, s/n: (b)(4).Smartablate generator (2 of 2), model # m-4900-07, s/n: (b)(4).(b)(4) are related to the same incident.
 
Event or Problem Description
It was reported that a patient underwent an ablation procedure for idiopathic ventricular tachycardia (idvt) with two (2) 8mm navigational bi-directional catheters and suffered a cardiac tamponade requiring pericardiocentesis.During the procedure, the smartablate generator would not display impedance for the two (2) 8mm navigational bi-directional catheters.The cables were replaced, the generator was re-booted, the generator was exchanged, and pace routing was confirmed to be disabled without success.The 8mm navigational bi-directional catheters were exchanged for a thermocool smarttouch uni-directional catheter.The smartablate generator initially did not display impedance for the smarttouch.The catheter connector was re-seated into the generator and the issue was resolved.During mapping with the smarttouch catheter, the patient became hypotensive and tamponade was confirmed via intracardiac echocardiogram.Pericardiocentesis yielded 150-200cc.The patient was reported to be in stable condition.Physician did not provide a causality opinion.Multiple attempts have been made to obtain clarification to this complaint.However, no further information has been made available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BIOSENSE WEBSTER, INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvacar
juarez, chihuahua 32599
MX  32599
MDR Report Key5938015
Report Number9673241-2016-00563
Device Sequence Number10930364
Product Code OAD
UDI-Device Identifier10846835003048
UDI-Public(01)10846835003048(11)151215(17)181130(10)17380030M
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P010068/S6
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Model NumberD-1260-04-S
Device Catalogue NumberBN7TCFJ8L
Device Lot Number17380030M
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2016
Initial Report FDA Received Date09/08/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
-
-