| Catalog Number |
AR-503-TTTE |
| Health Effect - Clinical Codes |
Pain (1994); Swelling (2091)
|
| Date of Event |
08/18/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to arthrex's follow-up requests.This is the first of four submissions from the same patient event.The others are 1220246-2016-00377 (cc95096-line 169385), 1220246-2016-00378 (cc95096-line 169386) and 1220246-2016-00379 (cc95096-line 169387).The device was not returned for evaluation and the contribution of the device to the reported event could not be determined.The information received regarding the event is not sufficient to determine the cause of the event.A dhr review for this catalog number, ar-503-ttte and lot number 108761229 combination found no related information.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2013 a patient underwent a left tka procedure.On (b)(6) 2016 the surgeon performed a revision left tka due to patient pain.During the revision procedure the surgeon explanted the tibial tray ar-503-ttte (lot 108761229)(cc95096-line 169384), femoral implant ar-503-pslf (lot 108761212) (cc95096-line 169385), bearing implant ar-503-be12 (lot 113601237) (cc95096-line 169386) and patella implant ar-504-psc9 (lot 113601237) (cc95096-line 169387).To complete the procedure the surgeon used another manufacturer's product.Patient was a female, dob: (b)(6).Medical records dated (b)(6) 2016 noted the following: patient having pain and a grinding sensation under kneecap when walking.Patient also noted the knee feels loose and unstable.Examination of the left knee with palpation demonstrated medial joint line and patellofemoral region tenderness but no lateral joint line tenderness.Trace effusion of the left knee was noted.Records noted patient has pain through a limited passive range of motion with crepitus and swelling.Medical records also noted x-rays showed periarticular osteophytes and joint space narrowing.
|
| |
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to arthrex's follow-up requests.This is the first of three follow-up submissions from the same patient event due to device evaluation.The others are 1220246-2016-00377f1 ((b)(4)) and 1220246-2016-00378f1 ((b)(4)).The information received regarding the event is not sufficient to determine the cause of the event.A dhr review for this catalog number, ar-503-ttte and lot number, 108761229 combination found no related information.The evaluation revealed that the returned device's critical mating features on surface a (surface that mates with the bearing) are within specification.The damages observed on the surfaces were most likely caused during the extraction of the implant.The contribution of the device to the reported event could not be determined.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2013 a patient underwent a left tka procedure.On 8/18/16 the surgeon performed a revision left tka due to patient pain.During the revision procedure the surgeon explanted the tibial tray ar-503-ttte (lot 108761229)((b)(4)), femoral implant ar-503-pslf (lot 108761212) ((b)(4)), bearing implant ar-503-be12 (lot 113601237) ((b)(4)) and patella implant ar-504-psc9 (lot 113601237) ((b)(4)).To complete the procedure the surgeon used another manufacturer's product.Patient was a female, dob (b)(6) 1961.Medical records dated 4/5/16 noted the following: patient having pain and a grinding sensation under kneecap when walking.Patient also noted the knee feels loose and unstable.Examination of the left knee with palpation demonstrated medial joint line and patellofemoral region tenderness but no lateral joint line tenderness.Trace effusion of the left knee was noted.Records noted patient has pain through a limited passive range of motion with crepitus and swelling.Medical records also noted x-rays showed periarticular osteophytes and joint space narrowing.
|
| |
|