MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0101
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/25/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: all available information was investigated and the reported inability to curve the steerable guiding catheter (sgc) was confirmed.A review of the lot history record revealed no manufacturing nonconformities.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.The investigation determined that the reported inability to curve the sgc was a result of an observed cable break; however, the investigation was unable to determine a conclusive cause for the sgc cable break.Additionally, the soft tip was noted to be torn.No torn material was noted to be missing.As it was reported that the clip did not get caught on the guide tip, there was no resistance noted during removal of the clip delivery system (cds) from the sgc and the patient anatomy is unlikely to have caused the visible damage, the torn soft tip may be a result of post-procedural handling; however, this cannot be definitively determined.The torn soft tip would not have contributed to the reported issues.Based on the information reviewed, there is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Event or Problem Description
This is filed based on returned device analysis which found that the steerable guiding catheter (sgc) soft tip was torn, which has the potential to cause or contribute to patient injury.It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 4.The steerable guiding catheter (sgc) was advanced without issue.While in the left atrium, the +/- knob was turned approximately 60 degrees in the + direction, but the tip did not respond.The sgc was removed without issue, and replaced.One clip was implanted, reducing the mr to 1.There were no adverse patient effects and no clinically significant delay in the procedure.Returned device analysis found that the sgc soft tip was torn.Follow up with the account confirmed that there was no resistance noted during removal of the delivery system from the sgc and the clip did not get caught on the sgc tip.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Common Device Name
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key5943244
Report Number2024168-2016-06021
Device Sequence Number1131985
Product Code DRA
UDI-Device Identifier08717648195921
UDI-Public(01)08717648195921(17)170430(10)60426U110
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K112239
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2017
Device Catalogue NumberSGC0101
Device Lot Number60426U110
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2016
Initial Report FDA Received Date09/12/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MITRACLIP SYSTEM
Patient Age81 YR
Patient Weight70
-
-