| Model Number |
393016 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
There is concern this device may have an early battery depletion.Previous data from follow-ups was reviewed and it was noted initially the implant date was entered correctly, but when the device was interrogated, it was noted the implant year was changed from 2015 to 2016 while maintaining the correct day and month.Also of note was from (b)(6) 2016: battery voltage had decreased from 39% to 9% @2.63 volts on (b)(6) 2016, voltage at implant is noted to have been 3.1 volts.Device has been programmed with normal voltage and pulse widths with acc and vcc features activated.A ram dump was obtained and the device was reset with no change in battery voltage or % remaining.Device remains implanted.Analysis of the data from the device was performed and device explant was recommended.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt, the icd interrogation revealed the battery status eri, three charging cycles were recorded to the device memory.The memory content of the device was inspected.The follow up data of august 30, 2016 12:12 pm documented an implantation date of (b)(6) 2016.The data further showed that the implantation date has been manually changed to (b)(6) 2015 at 12:19 pm on the same day.In a next step, the icd was subjected to an electrical analysis.During the analysis the current consumption of the icd was verified and found to be elevated.The ability of the device to deliver therapies was tested.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device delivered a defibrillation shock as specified, documenting a correct sensing and shock delivery.In particular, the specified energy level was reached.The device was opened to inspect the inner assembly and to check the functionality of the electronic module.Visually no deficiencies were observed.The current consumption of the electrical module was directly measured.The measurement confirmed an elevated current consumption.Further electrical investigations revealed that a low voltage capacitor of the electronic module was damaged, causing an increased current consumption.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process which might be related to the clinical observation.Particularly the final acceptance test proved the device functions to be fully functional.In conclusion, the clinical observation resulted from a damaged low voltage capacitor on the electronic module.There was no sign of any inconsistency during the manufacturing process, which might be related to the clinical observation.It is therefore assumed, that the damage occurred after the device shipment, however the date of occurrence was not determinable.Based on the analysis all therapy functions of the icd were available, while the device was implanted and in service.
|
| |
|
Event or Problem Description
|
|
There is concern this device may have an early battery depletion.Previous data from follow-ups was reviewed and it was noted initially the implant date was entered correctly, but when the device was interrogated, it was noted the implant year was changed from 2015 to 2016 while maintaining the correct day and month.Also, of note was from (b)(6) 2016: battery voltage had decreased from 39 percent to 9 percent @2.63 volts on (b)(6) 2016, voltage at implant is noted to have been 3.1 volts.Device has been programmed with normal voltage and pulse widths with acc and vcc features activated.A ram dump was obtained and the device was reset with no change in battery voltage or percentage remaining.Device remains implanted.Analysis of the data from the device was performed and device explant was recommended.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt, the icd interrogation revealed the battery status eri, three charging cycles were recorded to the device memory.The memory content of the device was inspected.The follow up data of august 30, 2016 12:12 pm documented an implantation date of(b)(6) 2016.The data further showed that the implantation date has been manually changed to (b)(6) 2015 at 12:19 pm on the same day.In a next step, the icd was subjected to an electrical analysis.During the analysis the current consumption of the icd was verified and found to be elevated.The ability of the device to deliver therapies was tested.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device delivered a defibrillation shock as specified, documenting a correct sensing and shock delivery.In particular, the specified energy level was reached.The device was opened to inspect the inner assembly and to check the functionality of the electronic module.Visually no deficiencies were observed.The current consumption of the electrical module was directly measured.The measurement confirmed an elevated current consumption.Further electrical investigations revealed that a low voltage capacitor of the electronic module was damaged, causing an increased current consumption.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process which might be related to the clinical observation.Particularly the final acceptance test proved the device functions to be fully functional.In conclusion, the clinical observation resulted from a damaged low voltage capacitor on the electronic module.There was no sign of any inconsistency during the manufacturing process, which might be related to the clinical observation.It is therefore assumed, that the damage occurred after the device shipment, however the date of occurrence was not determinable.Based on the analysis all therapy functions of the icd were available, while the device was implanted and in service.
|
| |
|
Event or Problem Description
|
|
There is concern this device may have an early battery depletion.Previous data from follow-ups was reviewed and it was noted initially the implant date was entered correctly, but when the device was interrogated, it was noted the implant year was changed from 2015 to 2016 while maintaining the correct day and month.Also of note was from may 24, 2016: battery voltage had decreased from 39% to 9% @2.63 volts on 08/30/2016, voltage at implant is noted to have been 3.1 volts.Device has been programmed with normal voltage and pulse widths with acc and vcc features activated.A ram dump was obtained and the device was reset with no change in battery voltage or % remaining.Device remains implanted.Analysis of the data from the device was performed and device explant was recommended.
|
| |
|
Search Alerts/Recalls
|