STRYKER INSTRUMENTS-KALAMAZOO CORDLESS DRIVER 4 HANDPIECE; INSTRUMENT, SURGICAL ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 4405000000 |
Device Problem
Failure to Shut Off (2939)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/24/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported event was duplicated.A service technician evaluated the device and observed that the trigger would stick, and run-on was observed.Upon disassembly, corrosion and wear were observed on the trigger assembly.The trigger assembly was replaced, preventative maintenance was performed, and the device was returned to the customer after passing final inspection.
|
|
Event Description
|
The device was returned to stryker instruments for service.During functional testing by a service technician, it was found that the reverse trigger was sticking and caused run on.No patient involvement, no delay, no medical intervention, and no adverse consequences were reported.
|
|