• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO CORDLESS DRIVER 4 HANDPIECE; INSTRUMENT, SURGICAL ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO CORDLESS DRIVER 4 HANDPIECE; INSTRUMENT, SURGICAL ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 4405000000
Device Problem Failure to Shut Off (2939)
Patient Problem No Patient Involvement (2645)
Event Date 08/24/2016
Event Type  malfunction  
Manufacturer Narrative
The reported event was duplicated.A service technician evaluated the device and observed that the trigger would stick, and run-on was observed.Upon disassembly, corrosion and wear were observed on the trigger assembly.The trigger assembly was replaced, preventative maintenance was performed, and the device was returned to the customer after passing final inspection.
 
Event Description
The device was returned to stryker instruments for service.During functional testing by a service technician, it was found that the reverse trigger was sticking and caused run on.No patient involvement, no delay, no medical intervention, and no adverse consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORDLESS DRIVER 4 HANDPIECE
Type of Device
INSTRUMENT, SURGICAL ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key5947839
MDR Text Key54628886
Report Number0001811755-2016-02200
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number4405000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2016
Date Manufacturer Received08/24/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-