| Model Number |
SPVA-2010 |
| Medical Device Problem Code |
Device Issue (2379)
|
| Health Effect - Clinical Code |
Hydrocephalus (3272)
|
| Date of Event |
06/13/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Sophysa has requested the user to send the device back to our facility for further analysis.Up until the day of this report (09/15/2016) , we are still waiting for the return of the device.
|
| |
|
Event or Problem Description
|
|
A patient suffered from a dilatation of the ventricular system after implantation of the polaris valve.No alteration of the distal catheter to the valve system and intracranial catheter was observed during the surgical act, however without operating the system prior to the implantation.
|
| |
|
Additional Manufacturer Narrative
|
|
The shape and mechanism of the spring inside the valve do not show any particular defect.After analyses of the valve, we could not find any possible explanation to this low pressure.The only clear non-conformity of this evaluation is the low pressure of the highest position, but this cannot explain the difficulties encountered by the user, as the valve was adjusted on the first and lowest position.In addition, we did not observe any obstruction of the valve, flow always remained possible.Obstruction is a known complication of a shunting system.It is described in our instructions for use.It can appear at any stage of the shunt, including for instance the catheters.We can however not make a clear affirmation as the rest of the shunting system has not been returned to us.
|
| |
|
Event or Problem Description
|
|
A patient suffered from a dilatation of the ventricular system after implantation of the polaris valve.No alteration of the distal catheter to the valve system and intracranial catheter was observed during the surgical act, however without operating the system prior to the implantation.
|
| |
|