|
On (b)(6) 2016, a (b)(6) female patient underwent evar for aaa.The patient was suitable for the procedure.The stent graft placement was conducted as planned; however, the confirmatory angiography confirmed massive type iv endoleak.A zenith flex aaa endovascular graft main body extension (esbe) was additionally placed inside the previously placed stent graft.The second confirmatory angiography confirmed that the type iv endoleak disappeared.There have been no adverse effects to the patient reported.
|
|
Investigation - evaluation: a review of complaint history, device history record, specifications, ifu, trends, and quality control data was conducted during the investigation.The complaint device was not returned; therefore no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that would contribute to this failure mode.There were no other reported complaints for this lot number.Based on the image review, the main body was introduced from the left and the contralateral gate positioned left anterior lateral.The contralateral gate was then accessed from the right as the ipsilateral limb was twisted counterclockwise to aid catheterization.The resulting anterior left gate position coupled with the posterior to anterior course of the right common iliac artery would have directed the right delivery system into the left main body wall between the second most distal and most distal stent bodies.It was at this location where the endoleak occurred.The leak was large and consistent with a fabric tear.Twist was evident between these two stent bodies and the flow divider stent.The leak was diminished by coverage from an esbe.The endoleak further lessened, but did not resolve, on follow up cta six days later.A type iii endoleak from a hole in the graft fabric much larger than suture hole was confirmed.The left anterior positioning of the eventual right gate necessitated counterclockwise twisting of the distal main body and pointed delivery advancement vectors straight at the hole's location.Although it cannot be determined whether this hole was preexisting or occurred during implantation, the possibility of tearing during right limb introduction was significantly increased.Significant findings relative to the use of the device were observed.The right / contralateral gate was deployed left anterior rather than right anterior.This necessitated distal main body twisting and would have directed advancement of the right limb deployment system into the main body wall where the tear occurred.Cause of adverse events was observed.The endoleak was secondary to a main body fabric tear.It was lessened by covering with an esbe.The ifu state, "main body placement - caution: to avoid any twist in the endovascular graft, during any rotation of the delivery system, be careful to rotate all of the components of the system together (from outer sheath to inner cannula)." based on the review, it can be seen that the right / contralateral gate was deployed left anterior rather than right anterior, which necessitated distal main body twisting.This would have directed the advancement of the right limb deployment system into the main body wall where the tear occurred.A large single tear was present in the left main body fabric between the second most distal and most distal main body stents.As the graft is fully inspected for damage during quality control and the tear occurred where the right limb deployment system would have been pushed forward, it is feasible to suggest that the damage to the graft occurred during implantation.Therefore, the root cause is due to user technique.Per the quality engineering risk assessment, no further action is warranted.
|