| Model Number |
4103202 |
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
08/22/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The sample was attached to 48" water head height which is approximately 1.73 psi (89.76mmhg), it did not leak at the umbrella valve.The sample was attached to the roller pump and tested at low speed and high speed.No leak was seen.Then the sample was tested at different pressure settings of negative and positive pressures, the valve leaked at positive pressure of 1310mmhg.These valves are designed for pressure relief at a negative pressure of -200mmhg and at a positive pressure <1300mmhg.The roller pump was ran to test for retrograde flow, the valve leaked at the pressure relief band as it is designed to do so.Evaluation in progress, not yet complete.
|
| |
|
Event or Problem Description
|
|
The foreign distributor (b)(6) reported an issue encountered with the use of the cardiopulmonary suction control device from one of their customers.The report stated that the device leaked for approximately 2-3 minutes during use.The report stated the device was removed from use and a different one was used to complete the procedure.The patient was given a transfusion to address the blood loss (volume not provided) with no patient complications reported as a result of the alleged issue.The device was returned to the manufacturer for evaluation.The device is a suction control device sold in bulk non-sterile form to the distributor for additional processing prior to end-use.
|
| |
|
Additional Manufacturer Narrative
|
|
No visual or functional anomalies were found with the device sample.The valve met specification.The device history record for the applicable lot of product did not show any manufacturing or quality concerns similar to the reported complaint condition.
|
| |
|
Search Alerts/Recalls
|