ANGIODYNAMICS EVLT KIT WITH SPOTLIGHT OPS; ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number H787EVLT80OPS5
Medical Device Problem Codes Crack (1135); Fracture (1260); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/15/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
As reported (b)(6) 2016, a (b)(6), male patient presented for an endovenous laser procedure.During preparation for the procedure, when the laser fiber was removed from the sterile packaging, it was noted the fiber was fractured.The device was set aside and a new of the same device was used to successfully complete the procedure.There was no harm or injury to the patient due to the event as the device did not come into contact with the patient.The reported disposable device has been returned to the manufacturer for a device evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was an 80 ops fiber.A visual inspection noted that 67.7cm from the gripper, the fiber is fractured.The customers reported complaint of a broken fiber is confirmed.The patient was unaffected due to this event.A review of the lot history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.At the vendor facility, during the manufacturing process, fibers are repeatedly bent and coiled and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to shipment.During the packaging step, the fibers are inspected for damage.It is unlikely that the fiber was fractured prior to packaging.The most likely root cause to this event is due to handling after the device left the angiodynamics facility.He directions for use which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVLT KIT WITH SPOTLIGHT OPS
Common Device Name
ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
MDR Report Key5966063
Report Number1319211-2016-00130
Device Sequence Number93414
Product Code GEX
UDI-Device IdentifierH787EVLT80OPS5
UDI-PublicH787EVLT80OPS5
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K100199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date02/28/2019
Device Catalogue NumberH787EVLT80OPS5
Device Lot Number4994817
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/02/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 08/31/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date09/21/2016
Supplement Report FDA Received Date03/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age40 YR
-
-