| Model Number |
HJL016-40-N |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Physical Property Issue (3008)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
08/23/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation but since it had been prepped for use and exposed to blood/preservatives we were unable to reach any conclusions from our investigation.The device history record for this lot was reviewed.The lot passed all inspection criteria including: visual inspection, external diameter, and leak testing at simulated physiological pressure.This is a bioprosthetic device and variation along the graft is normal.Our ifu does state that a "small amount of oozing is acceptable".At this time we are unable to determine the cause of the incident but believe it was an isolated event.
|
| |
|
Event or Problem Description
|
|
During implantation of the graft, the surgeon pressurized it and noticed that a one inch section near the end appeared thin, lighter in color and started to slightly leak.The doctor removed the graft and used a different graft in stock without further incident.
|
| |
|