| Model Number |
85451 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/30/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
On completion of the investigation, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The device was placed into a 6fr ansel sheath hub and the stent slipped back onto the balloon catheter before it could be placed in the patient.The patient wasn't harmed.
|
| |
|
Additional Manufacturer Narrative
|
|
Engineering analysis: the icast was removed from the packaging and inspected to determine the cause of the complaint.The delivery system was decontaminated and inspected for damage.There was no damage seen.The stent was shifted on the balloon distally.The proximal end of the stent was slightly flared from being dislodged.The stent diameter was measured and was 2.13mm.This diameter is indicative of the 59 stents measured during the quality performance testing inspection.The returned device balloon surface was evaluated to determine if the stent was crimped properly during manufacturing, when the stent is crimped on to the folded balloon the stent frame leaves impressions of the stent frame on the surface of the balloon.The impressions indicate if the stent was properly crimped on to the balloon.The crimped stent impressions were clearly visible indicating that the stent was properly crimped during manufacturing.During the final lot qualification data shows that all 59 test samples were able to pass through the 6fr introducer sheath without issue.Since december of 2013 well over (b)(4) test units have been passed through the introducer sheath without a failure for the inability of the stent to pass through the introducer sheath.The product in question has also been subjected to simulated use in a tortuous iliac artery model whereas the stent delivery system is advanced contra laterally over the iliac arch through a 6fr 55cm long cook check flow performer introducer sheath.The stent is then deployed at nominal pressure as specified on the product label and the balloon deflated and withdrawn back through the introducer sheath.This testing was conducted numerous times while being submerged in a heated water bath at 37°c (body temperature) during design verification testing of the product.The introducer sheath used in the case was not returned.Engineering summary: a full review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atriums final lot qualification testing.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath; ability to deploy the stent at nominal pressure (8atm); ability to withdraw the deflated balloon catheter back through the labeled introducer sheath; ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures; balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm); manifold to shaft tensile testing.Samples when tensile tested must not break or separate below 3.37lbs.Result: all 59 quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing related to the stent.None of the 59 samples tested had any stent movement while passing the device through the introducer sheath.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the device and or lot of stent delivery systems in question.Clinical evaluation: there are several possibilities that can cause stent dislodgement including but not limited to manipulation of the stent prior to use, overuse of the sheath causing failure of the hemostatic valve or a hemostatic valve that is too tight.The instructions for use state that, "special care must be taken not to handle or in any way disrupt the placement of the icast stent on the balloon." insufficient stent securement during access through the sheath may lead to the need for removal of the stent delivery system.
|
| |
|
Search Alerts/Recalls
|