Catalog Number 1235250-18
Health Effect - Clinical Code Perforation (2001)
Date of Event 09/02/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of perforation, as listed in the absorb gt1 instructions for use (ifu) is a known adverse event associated with the use of a coronary scaffold in native coronary arteries.It should be noted that the ifu states: do not exceed rbp (rated burst pressure) of the balloon or maximum deployment diameter of the scaffold.Based on the case information and related record review, a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that during the procedure to treat a mildly calcified lesion in the heavily tortuous mid to distal left anterior descending coronary artery, a 2.5mm x 18mm absorb gt1 scaffold delivery system was advanced with ease to the target lesion.When deploying the gt1 scaffold without issue at 18 atmospheres, a distal edge perforation occurred.The perforation was successfully sealed via prolonged inflations with an unspecified 2.25mm balloon.There were no device issues, no adverse patient sequelae, and no occurrence of a clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Common Device Name
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key5976007
Report Number2024168-2016-06414
Device Sequence Number10871470
Product Code PNY
UDI-Device Identifier08717648212963
UDI-Public(01)08717648212963(17)170509(10)6051061
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P150023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/09/2017
Device Catalogue Number1235250-18
Device Lot Number6051061
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2016
Initial Report FDA Received Date09/24/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-