| Model Number |
ZLB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
08/31/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zlb00 19.5 diopter intraocular lens (iol) was explanted due to unexpected post-op refraction.At explant there was no incision enlargement, no vitrectomy and no sutures were required.The replacement lens was another zlb00 of a higher (20.5).There was no post-op patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device was not returned to the manufacturer for evaluation.Device inspection could not be performed, therefore the reported complaint could not be verified.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of a product deficiency.All pertinent information available to the manufacturer has been submitted.
|
| |
|