|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The evaluation of the photo provided from the customer shows incorrect orientation of a sphincterotome in the endoscopic view.However, it is unknown which device (either device 1, lot w3717533, mdr report number 1037905-2016-00379, or device 2, lot w3748130, which is this report) is pictured.Without return of the complaint device, further investigation cannot be completed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Improper cutting wire orientation can occur if the distal end of the catheter is shaped manually.This sphincterotome catheter is pre-curved and is provided with a pre-curved stylet in the distal tip of the catheter.This obviates the need for manual formation.The instructions for use contain the following comment: ¿note: do not apply manual pressure to tip or cutting wire of the sphincterotome to influence orientation, as this may result in damage to device.¿ other factors that can contribute to improper cutting wire orientation include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use state: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip.Note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable.¿ the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all tri-tome pc triple lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
As reported to customer relations: "during an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook tri-tome pc triple lumen sphincterotome.As they bowed the sphincterotome, it bowed to the left [incorrect cutting wire orientation] (see mdr report number 1037905-2016-00379).It was removed and they used another sphincterotome which had the same results [second device also had incorrect cutting wire orientation].They used a different cook sphincterotome to continue the procedure with no further complications.There was no harm to the patient.".
|