| Catalog Number |
9734403 |
| Medical Device Problem Codes |
Bent (1059); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/30/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not made available from the site.Device lot number not available.Device manufacturing date is dependent on lot number, therefore, unavailable.Return requested for suspect thoracic probe.No parts have been received by manufacturer for analysis.No further issues have been reported.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that, while in a posterior spinal fusion procedure, the site's thoracic probe was bent.No further details regarding the damage, or how it occurred, were provided.The surgeon opted to continue and completed the procedure with the use of the navigation system.There was no delay of therapy.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of returned suspect thoracic probe was unable to replicate the reported issue.The tip of the probe does not appear to be bent.There are nicks along the edge at the tip, but no bend.There are also several small impact marks at the back end but they do not effect function.No problem found.The reported event could not be duplicated by medtronic personnel.
|
| |
|