(b)(4).Age at time of event: 18 years or older.Device is combination product.(b)(4).Device evaluated by mfr: the stent delivery system (sds) was returned for analysis.A visual and tactile examination found multiple kinking on the hypotube shaft and the shaft itself was broken at 200mm from the distal end of the hub.A visual and tactile examination of the outer and mid-shaft section found no issues with their profile.The outer extrusion, inner lumen and bi-component bond showed no signs of damage or strain.A visual examination of the crimped stent found no issues with the crimped stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The balloon cone profiles were reviewed and no issues were noted with the overall balloon profile.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.The bumper tip of the device showed no signs of damage.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
Reportable based on device analysis completed on 01-sep-2016.It was reported that hub damage and crossing difficulties were encountered.The non totally occluded target lesion was located in a coronary artery.A 2.50x24mm promus element ¿ drug-eluting stent was advanced to treat the lesion.However, significant resistance was encountered and the hub was damaged while negotiating the device.The procedure was completed with another 2.25x24 promus element ¿ drug-eluting stent.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed hypotube break.
|