Catalog Number 82-3112
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Information (3190)
Event Date 08/12/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Upon completion of the investigation a follow up report will be filed.
 
Event Description
Rotator dislokation.
 
Manufacturer Narrative
(b)(4).Upon completion of the investigation it was noted that the images were taken of the ¿as received¿ valve.The position of the cam when valve was received was 110mmh2o.The valve was hydrated for 24 hours.The valve was visually inspected: no defects were noted.The valve was tested for programming.With programmer 82-3126 with serial number (b)(4), the valve passed the test.The valve was flushed, the valve passed the test no occlusion was noted.The valve was leak tested, only leaked from the needle holes in the needle chamber.The catheters were irrigated, no occlusions noted.The valve was reflux tested, the valve passed the test.The valve was dried.The valve was then pressure tested, the valve passed the test.Review of the history device records confirmed the valve product code 82-3112 with lot clmc0w, conformed to the specifications when released to stock in 17th november 2010.No root cause could be determined, as the problem reported by the customer could not be duplicated.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAKIM PROGRAMMABLE MICRO VALVE
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
rue girardet 29
ch-2400 le locle
SZ 
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key5983312
MDR Text Key55813770
Report Number1226348-2016-10687
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K980778
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2015
Device Catalogue Number82-3112
Device Lot NumberCLMC0W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/04/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age5 YR
-
-