It was reported that a cleo 90 infusion set adhesive was stuck to the lid.Blood glucose levels were adversely affected and reported at 600mg/dl the patient changed supplies and then delivered bolus to resolve the high blood glucose.It was reported that the blood glucose was under control.No permanent injury was reported.See mfr: 2183502-2016-02095, 2183502-2016-02096, 2183502-2016-02097, 2183502-2016-02098, 2183502-2016-02099, 2183502-2016-02100, 2183502-2016-02101, 2183502-2016-02103, 2183502-2016-02104, 2183502-2016-02115, 2183502-2016-02116, 2183502-2016-02117, 2183502-2016-02118, 2183502-2016-02119, 2183502-2016-02120, 2183502-2016-02121, 2183502-2016-02122, 2183502-2016-02123, and 2183502-2016-02124.
|