• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125350-33
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/13/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.Visual inspections were performed on the returned device.The tip damage was confirmed.The investigation was unable to determine a conclusive cause for the reported difficulties.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the 3.5 x 33 mm xience alpine stent delivery system was removed from the packaging and the protective sheath was removed without difficulty.It was noted that the orange tip of the stent delivery system was damaged.The device was not used and there was no patient involvement.A second xience alpine stent delivery system was used without issue.There was no adverse patient effect or clinically significant delay.Upon return, device analysis noted that the tip of the device was stretched and torn.Follow up with the site indicates the damage noted is the reported tip damage.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key5988800
Report Number2024168-2016-06540
Device Sequence Number19007130
Product Code NIQ
UDI-Device Identifier08717648199561
UDI-Public(01)08717648199561(17)190330(10)6032341
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2019
Device Catalogue Number1125350-33
Device Lot Number6032341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2016
Initial Report FDA Received Date09/29/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-