|
Summary: the echelon 10 microcatheter and the rotating hemostatic valve (rhv) were not returned.As viewed through the returned packaging, unreported coil severing and stretching was found.Unreported complete separation of the device positioning unit (dpu) and the introducer sheath was found.Unreported protrusion of the severed coil was found located approximately 41.0 centimeters off the distal tip of the green introducer.External force caused a ductile fracture resulting in the coil severing into two pieces.No material defects were found.There is no mechanical sheath damage at the coil protrusion site.The remainder of the coil past the protrusion site is undamaged.Located on the top proximal end of the resheathing tool in the open cutout section, the v notch has been severely damaged with the damaged edges elevated above the surface plane.The locking mechanism has indentation, compression, and stretching damage.No manufacturing defects were found.Due to the severe damage and without the return of the microcatheter used in the procedure, no duplication of the field complaint could be performed.The circumstances of how and when all the above unreported damage occurred to the unit cannot be determined as it was reported, ¿¿there were no damages to the device when it was removed from the patient¿¿ conclusion: the complaint of the coil becoming stuck inside the proximal end of the microcatheter is not confirmed.Due to all the unreported damage to the microcoil system, the severing of the coil, and without the return of the microcatheter used in the procedure, the root cause of the coil becoming stuck cannot be determined, however the evidence as returned highly suggests that the most likely contributing factor to the coil becoming ¿stuck¿ inside the proximal end of the microcatheter may have occurred when the microcoil system was first unlocked for use and the sheath was retracted straight back instead of up at a forty-five degree angle and then back.When the sheath was pulled straight back, the locking mechanism caught the inside of the v notch of the resheathing tool and became embedded.In addition, the locking mechanism may not have been fully disengaged off the core wire.The sheath also caught the v notches extended edges.This produced a binding action between the device positioning unit (dpu), the sheath, and the coil.This binding action produced significant resistance which prevented the coil from advancing past the proximal end of the microcatheter and may have produced a portion of the unreported coil damage found.For optimum product performance and to prevent potential complications, the instructions for use (ifu) recommends, ¿hold the introducer sheath (loosely-looped) in the left hand.Keeping the introducer tip near the re-sheathing tool, grasp the distal end of the re-sheathing tool between your left thumb and forefinger.Grasp the clear tab near the end of the introducer sheath body with the thumb and forefinger of your other hand.Gently pull the clear tab of the introducer sheath out and away from the re sheathing tool at a 45-degree angle to unlock the microcoil.Continue to pull the tab until an additional 0.5 to 1.0 inches (1.3 to 2.5 cm) of the translucent material is exposed.Gently fold the translucent tab towards the distal end, and firmly grasp the distal end of the re sheathing tool and the translucent tab between your thumb and forefinger, as shown in figure 3¿¿ caution: if unusual friction is noticed during advancement or retraction of the microcoil system, verify the locking mechanism, or clear tab is unlocked and pulled out from the resheathing tool approximately 1in.(2-3cm).¿ in addition, without the return the echelon 10 microcatheter and the rhv used in the procedure, it cannot be determined if these components had any additional contributions to the complaint event.Dhr: a review of the manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Based on the information and the analysis, there were secondary damages that were not initially reported however procedural factors outlined in the ifu likely contributed to the damages.Without return of concomitant products it cannot be determined if those components had any additional contributions to the complaint event.Additionally, review of this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.Therefore, no corrective actions will be taken at this time.(b)(4).
|