DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.110_LL |
Device Problems
Sticking (1597); Component Missing (2306)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/20/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during service and repair/pre-testing, it was observed that the reverse trigger on the small battery drive device was sticking, the device would not run and one of the discs was missing.The event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device would not run, the triggers were sticky, and one of the safety discs was missing, the electronic control unite was damaged, the triggers were worn and the internal components were corroded.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to worn components from normal use and servicing over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|