• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.110_LL
Device Problems Sticking (1597); Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 09/20/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during service and repair/pre-testing, it was observed that the reverse trigger on the small battery drive device was sticking, the device would not run and one of the discs was missing.The event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Additional narrative: the actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device would not run, the triggers were sticky, and one of the safety discs was missing, the electronic control unite was damaged, the triggers were worn and the internal components were corroded.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to worn components from normal use and servicing over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ  CH4437
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
hauptstrasse 24
waldenburg CH443 7
SZ   CH4437
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5994813
MDR Text Key56585539
Report Number8030965-2016-15046
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Service and Testing Personnel
Type of Report Initial,Followup
Report Date 09/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.110_LL
Device Lot Number1368
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received10/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-