| Model Number |
5214402 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Paresis (1998)
|
| Date of Event |
09/05/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No device has been returned for evaluation.It is unknown if patient has neurologic condition which may have affected neuromonitoring.Radiographs were received confirming the malplacement of the initial implant.Action to correct malplacement included placement of a smaller interbody device.The root cause of this reported event is unknown; however, review of the radiographs and operational notes indicated suboptimal interbody placement related to implant size selection or anatomical characteristics of the patient.There is no plan for revision, and the patient will continue to be monitored to determine if the paresis condition resolves.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016 an interbody device was inserted at l4l5.Intra-operative lateral and ap radiographs showed the device was lateral to the disc space and required adjustment.The cage was replaced with a smaller interbody implant and the case was completed without further difficulties.On 09/06/2016 a neural motor deficit of the patient's left foot was reported.
|
| |
|