| Catalog Number |
700-7177-001 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
09/16/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer (fse) was dispatched to the site to evaluate the instrument on 09/16/2016.The fse found a crack in the tubing from the color gravity module (cgm) to the probe.The fse replaced the tubing from the cgm to the probe to resolve the leak.The beckman coulter internal identifier for this report is (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer stated that there was a contained fluid leak of approximately 7ml from the probe on an ichemvelocity automated urine chemistry system.The customer was wearing personal protective equipment (ppe) consisting of gloves at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no injury or effect to patient treatment related to this event.
|
| |
|