The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Sorin group (b)(4) manufactures the sorin heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the sorin heater-cooler system 3t displayed an error code on the patient side during a procedure.There was no report of patient injury.A sorin group field service representative was dispatched to the facility to investigate.The service representative confirmed the reported error code and identified that the stirrer on the patient side had difficulty moving.The stirrer was freed and all cable connections between the distribution board and processor board were checked.Subsequent functional testing was performed without any issue and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.No trend has been identified for this type of issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.Evaluated on site by sorin service rep.
|