This is a supplemental report for mfr report # 8010047-2016-01269 to provide additional information based on the evaluation of the device.Device manufacturer date was identified and filled in the report.The subject device was returned to olympus medical systems corp (omsc) for evaluation.As a result of evaluation of omsc on october 5, 2016, omsc confirmed that the ptfe pad of the device was partially peeled.The manufacturing record was reviewed with no irregularities.This type of ptfe pad damage is most likely related to the operator's technique.Based on the past similar cases, it was known that the ptfe pad was partially peeled when the user continued to activate seal & cut mode without contacting tissue (including after the tissue already cut).The instruction manual of the subject device already states; warnings: do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the ptfe pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.
|