AMEDA, INC. PURELY YOURS; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem Burn(s) (1757)
Event Date 09/16/2016
Event Type  Injury  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance, burning, melting or charring were observed.The returned batteries were observably damaged.Additional testing confirmed that batteries may leak when the upper middle battery is placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2016 to report pumping with the purely yours breast pump on this same date using battery power.She had her hands underneath the pump base as she expressed her milk.She reports hearing a gurgle sound from the base and felt hot fluid on her right hand.She noted the fluid was leaking from the battery compartment, and was dark.She stopped pumping, threw away the breast milk she just expressed and noted a small red burn between her small and ring fingers near the palm of her hand.The area was red, slightly swollen but not blistered.She ran her hand under cold water and did not seek medical attention for the burn.Customer later removed the door of the battery compartment, finding some batteries had exploded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key6002906
MDR Text Key56584134
Report Number3009974348-2016-00228
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 09/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received09/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age29 YR
-
-