• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY IMPLANTS N.V. SURGIGUIDE; SYSTEM, IMAGE PROCESSING, RADIOLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY IMPLANTS N.V. SURGIGUIDE; SYSTEM, IMAGE PROCESSING, RADIOLOGICAL Back to Search Results
Catalog Number 37503
Device Problem Positioning Problem (3009)
Patient Problem Failure of Implant (1924)
Event Date 02/29/2016
Event Type  Injury  
Manufacturer Narrative
Since this event resulted in medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the malfunction would be likely to cause or contribute to a serious injury should it recur.As such, this event meets the definition of a reportable event per 21 cfr part 803.Evaluation found there is scatter in the images and no clear outline of the mucosa information.There is nothing wrong with the fit on the plaster model or with the design in the planning file.
 
Event Description
It was reported that a surgiguide was used to create the osteotomy for three dental implants (tooth position 5, 6 and 7).After the first drilling step the surgical guide was removed to complete the additional drilling steps free handed according to the pilot guide drilling protocol.After placing the implants free handed, the customer discovered that the three implants were positioned too buccal.A second surgery is needed to correct the positioning of the implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIGUIDE
Type of Device
SYSTEM, IMAGE PROCESSING, RADIOLOGICAL
Manufacturer (Section D)
DENTSPLY IMPLANTS N.V.
research campus 10
hasselt limburg, B-350 0
BE  B-3500
Manufacturer (Section G)
DENTSPLY IMPLANTS N.V.
research campus 10
hasselt limburg, B-350 0
BE   B-3500
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6003092
MDR Text Key56590876
Report Number3007362683-2016-00012
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 08/29/2016
3 Devices were Involved in the Event: 1   2   3  
1 Patient was Involved in the Event
Date FDA Received10/06/2016
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number37503
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/29/2016
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-