• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH ENDOSTAT 3 BIPOLAR/MONOPOLAR ELECTROSURGICAL GENERATOR; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH ENDOSTAT 3 BIPOLAR/MONOPOLAR ELECTROSURGICAL GENERATOR; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00540200
Device Problems Failure to Deliver Energy (1211); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/16/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4) relates to (b)(4) for the reported event of no output in control cut and monopolar mode.(b)(4) for the reported event of generator displays system fault error.The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that an endostat iii rf generator was used during a colonoscopy procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the endostat has no output in control cut and monopolar mode.The analyzer prompts an error message that the rf detached.The mode was switch to cut only and the procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Investigation results: analysis of the returned endostat iii rf generator revealed that the unit appears to be in good physical condition and the water bottle was not returned in the box.Functional testing was performed and it passed all the requirements.The complaint that the device ¿had no output and was having an error message¿ could not be confirmed.The most probable root cause classification of not confirmed was assigned for this event since the device operated with no issues.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database confirmed that no similar complaints exist for this generator.
 
Event Description
It was reported to boston scientific corporation that an endostat iii rf generator was used during a colonoscopy procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the endostat has no output in control cut and monopolar mode.The analyzer prompts an error message that the rf detached.The mode was switch to cut only and the procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSTAT 3 BIPOLAR/MONOPOLAR ELECTROSURGICAL GENERATOR
Type of Device
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
MEDICAL SCIENTIFIC
125 john hancock road
taunton MA 02780
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key6003897
MDR Text Key56928737
Report Number3005099803-2016-02950
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K913881
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00540200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/16/2016
Initial Date FDA Received10/06/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/08/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/17/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-