Model Number M00540200 |
Device Problems
Failure to Deliver Energy (1211); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/16/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) relates to (b)(4) for the reported event of no output in control cut and monopolar mode.(b)(4) for the reported event of generator displays system fault error.The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an endostat iii rf generator was used during a colonoscopy procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the endostat has no output in control cut and monopolar mode.The analyzer prompts an error message that the rf detached.The mode was switch to cut only and the procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Investigation results: analysis of the returned endostat iii rf generator revealed that the unit appears to be in good physical condition and the water bottle was not returned in the box.Functional testing was performed and it passed all the requirements.The complaint that the device ¿had no output and was having an error message¿ could not be confirmed.The most probable root cause classification of not confirmed was assigned for this event since the device operated with no issues.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database confirmed that no similar complaints exist for this generator.
|
|
Event Description
|
It was reported to boston scientific corporation that an endostat iii rf generator was used during a colonoscopy procedure performed on (b)(6) 2016.According to the complainant, during the procedure, the endostat has no output in control cut and monopolar mode.The analyzer prompts an error message that the rf detached.The mode was switch to cut only and the procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|