• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2290
Device Problems Under-Sensing (1661); Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2016
Event Type  Injury  
Manufacturer Narrative
Product event summary: it could not be confirmed that sensing from the analyzer was variable, it passed all functional and systems tests.It was found that the analyzer's patient connector had disturbed pins.
 
Event Description
It was reported that during an implant procedure, sensing from the analyzer was variable.The lead, cable, and adapters on the programmer were changed without effect.The programmer was exchanged, and there were much better values when using a new programmer and analyzer.It was noted that the analyzer appeared to have been reporting r waves below their actual value.The right ventricular (rv) lead remains in use, although it was repositioned multiple times.The analyzer has been returned to service.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6005149
MDR Text Key56765452
Report Number2182208-2016-02251
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2290
Device Catalogue Number2290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/02/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-