• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 ARCTICSUN GEL PADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 ARCTICSUN GEL PADS Back to Search Results
Catalog Number 317-05-02
Device Problems Air Leak (1008); Restricted Flow rate (1248)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/01/2016
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿attach the pad¿s line connectors to the patient line manifolds.Begin circulating water through the pads using either patient temperature control mode (automatic) or water temperature control mode (manual).If the pads fail to prime or a significant continuous air leak is observed in the pad return line, check connections, then if needed replace the leaking pad.Once the pads are primed, assure the flow rate displayed on the control panel is greater than 2.3 liters per minute, which is the minimum flow rate for a full pad kit.¿ (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the pads attracted air in the connection places, allegedly causing low-flow.However, the patient was kept stable, so therapy was continued with a replacement of the pads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTICSUN GEL PADS
Type of Device
ARCTICSUN GEL PADS
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6010671
MDR Text Key57319892
Report Number1018233-2016-01363
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
PMA/PMN Number
K142702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2017
Device Catalogue Number317-05-02
Device Lot NumberNGAQ5605
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/13/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-