• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIS INTERNATIONAL ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM; AUTOMATED URINALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIS INTERNATIONAL ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM; AUTOMATED URINALYSIS SYSTEM Back to Search Results
Catalog Number 700-7177-001
Device Problem Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/16/2016
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument.The fse found that a strip was stuck and was blocking the strip reader module (srm).The fse removed the strip to resolve the reported issue.(b)(4).
 
Event Description
The customer reported the ichem velocity automated urine chemistry system was failing ca controls for bilirubin parameter.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.
 
Manufacturer Narrative
The device date of manufacture was changed from 06/07/2013 to 03/14/2014.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM
Type of Device
AUTOMATED URINALYSIS SYSTEM
Manufacturer (Section D)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer (Section G)
IRIS INTERNATIONAL
9172 eton ave
chatsworth CA 91311
Manufacturer Contact
gopal mohanty
9172 eton avenue
chatsworth, ca, CA 91311
8185277379
MDR Report Key6010958
MDR Text Key56918276
Report Number2023446-2016-00383
Device Sequence Number1
Product Code KQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101852
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number700-7177-001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-