The reason for this revision surgery was an infection.The previous surgery and the revision detailed in this investigation occurred 64 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), and product complaint report (pcr) database shows that the reported components used in the previous surgery met design and manufacturing requirements.There was one non-conforming material report (ncmr) ncmr, (b)(4), associated with the product for reject code o8 (improper packaging) disposition-rework (40), units 1-40 were reworked ((b)(4)) and found acceptable after rework.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
|