MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FOUNDATION PRIMARY KNEE; INSERT, STD., PRIMARY, SIZE4 11MM CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FOUNDATION PRIMARY KNEE; INSERT, STD., PRIMARY, SIZE4 11MM CM Back to Search Results
Catalog Number 385-11-504
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/15/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having an infection.
 
Additional Manufacturer Narrative
The reason for this revision surgery was the patient had an infection.The previous surgery and the revision detailed in this investigation occurred over 8.6 yrs.Apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) and product complaint report (pcr) database records show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION PRIMARY KNEE
Common Device Name
INSERT, STD., PRIMARY, SIZE4 11MM CM
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6011722
Report Number1644408-2016-00745
Device Sequence Number18842278
Product Code JWH
UDI-Device Identifier00888912117838
UDI-Public(01)00888912117838
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K000590
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Medical Equipment Company Technician/Representative
Device Expiration Date06/28/2008
Device Catalogue Number385-11-504
Device Lot Number890931
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 09/15/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/07/2016
Supplement Report FDA Received Date11/11/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-