| Catalog Number |
6706366 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
09/20/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer (fse) evaluated the instrument at the customer's site.The fse found that the tubing through valve lv8 was defective, causing the leak.The fse replaced the tubing, resolving the leak.The repairs were verified per established service procedures.The beckman coulter internal identifier for this report is (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported less than 1 ml of uncontained fluid leaked from the probe in a coulter ac t diff analyzer onto the counter during normal instrument operation.There were vacuum error messages reported by the customer at the time of the event.The customer was wearing personal protective equipment (ppe) consisting of laboratory coat, goggles and gloves at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection to the event.There was no impact to patient results and controls.
|
| |
|