| Catalog Number |
48811234 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/13/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cervical plate was being used.The locking nut was engaged on the plate causing the spring bar to be damaged upon screw insertion.Another 34mm plate was used.Screw is still stuck in plate.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, device history review, complaint history review, risk assessment; result: the device was confirmed visually to have a deformed spring bar.Manufacturing files were reviewed and no anomalies were found.Conclusion: the cause is not using the recommended drill guides as per the stg.
|
| |
|
Event or Problem Description
|
|
It was reported; that the cervical plate was being used.The locking nut was engaged on the plate causing the spring bar to be damaged upon screw insertion.Another 34mm plate was used.Screw is still stuck in plate.
|
| |
|