SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP CA 19-9 ASSAY; IMMUNOLOGICAL TEST FOR CA 19-9 ANTIGEN
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/13/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The cause for the falsely elevated advia centaur xp ca 19-9 result is unknown.The customer's quality control results were all acceptable at the time of the event.Siemens is investigating.The limitations section of the instructions for use (ifu) states: "warning": "do not use the advia centaur ca 19-9 assay as a screening test or for diagnosis.Do not predict disease recurrence solely on levels of advia centaur ca 19-9.Normal levels of advia centaur ca 19-9 do not always preclude the presence of disease." "note" : "do not interpret serum levels of ca 19-9 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed gi carcinoma frequently have levels of ca 19-9 within the range observed in healthy individuals.Additionally, elevated levels of ca 19-9 can be observed in patients with nonmalignant diseases.Measurements of ca 19-9 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation." "the concentration of ca 19-9 in a given specimen determined with assays from different manufacturers can vary because of differences in assay methods, calibration, and reagent specificity.Ca 19-9 determined with different manufacturers' assays will vary depending on the method of standardization and antibody specificity.Therefore, it is important to use assay specific values to evaluate quality control results.".
|
|
Event Description
|
A falsely elevated advia centaur xp ca 19-9 result was obtained by the customer, and the result was questioned by the physician.The patient had no clinical reason to support an elevated advia centaur ca 19-9 result.Repeated testing was performed on an alternate ca 19-9 test method at an alternate laboratory and the result was lower.The patient was redrawn, and ca 19-9 testing was performed on the advia centaur xp and alternate test method.The ca 19-9 result was elevated on the advia centaur xp and lower on the alternate test method.There are no reports that treatment was altered or prescribed or adverse health consequences due to the discordant advia centaur xp ca 19-9 result, however, the patient experienced a high level of anxiety due to the elevated result.
|
|
Manufacturer Narrative
|
Siemens filed the initial mdr 1219913-2016-00179 on 10/11/2016 for a falsely elevated advia centaur xp ca 19-9 patient result.On 11/02/2016 - additional information: the cause for the falsely elevated advia centaur xp ca 19-9 patient result compared to a lower alternate ca 19-9 test method result is unknown.The patient's sample is not available for further investigation.Immunoassays are also subject to a number of interferences including those caused by endogenous antibodies.Interference can occur because of heterophile antibodies, anti-animal antibodies and auto antibodies.Patients exposed to animals or animal serum products can be prone to this interference and anomalous values may be observed.The interfering antibodies can give rise to a falsely high or less commonly a falsely low result.The erroneous result is recognized as being inconsistent with the patient's clinical picture.The laboratory procedures generally used to identify the presence of interfering antibody are serial dilutions to demonstrate a nonlinear response.Please note the interferent may not necessarily be due to an interfering antibody but may be due to other exogenous interferences such as drugs, nutritional supplements and/or herbal medicine in the blood.No conclusion can be drawn.The warning section of the instructions for use (ifu) states: "the concentration of ca 19-9 in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity.The results reported by the laboratory to the physician must include the identity of the assay for ca 19-9 used.Values obtained with different assay methods cannot be used interchangeably.If, in the course of monitoring a patient, the assay method used for determining serial levels of ca 19-9 is changed, the laboratory must perform additional serial testing to confirm baseline values.The advia centaur ca 19-9 assay is based on the 1116-ns-19-9 antibody available through agreement with fujirebio diagnostics, inc.Assays using antibodies other than 1116-ns-19-9 may give different results." the limitations section of the instructions for use (ifu) states the following: "warning." "do not use the advia centaur ca 19-9 assay as a screening test or for diagnosis.Do not predict disease recurrence solely on levels of advia centaur ca 19-9.Normal levels of advia centaur ca 19-9 do not always preclude the presence of disease." "note:" "do not interpret serum levels of ca 19-9 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed gi carcinoma frequently have levels of ca 19-9 within the range observed in healthy individuals.Additionally, elevated levels of ca 19-9 can be observed in patients with nonmalignant diseases.Measurements of ca 19-9 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation." "the concentration of ca 19-9 in a given specimen determined with assays from different manufacturers can vary because of differences in assay methods, calibration, and reagent specificity.Ca 19-9 determined with different manufacturers' assays will vary depending on the method of standardization and antibody specificity.Therefore, it is important to use assay specific values to evaluate quality control results." "heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays.Patients routinely exposed to animals or to animal serum products can be prone to this interference and anomalous values may be observed.Additional information may be required for diagnosis." the instrument is performing within specifications.No further investigation is required.
|
|
Search Alerts/Recalls
|
|
|