It was reported that during a second and third pip fusion case, the ar-4157-24 retrofusion screw, 24mm broke while attempting to insert in the second pip joint.Bony prep was followed per surgical technique.The screw was advanced in the proximal phalanx and was unable to advance further than 2/3 of proximal threads being in the im canal.K wire was taken circumferentially around screw in attempts to soften surrounding bone and remove screw but was unsuccessful.Since screw could not be removed, attempts were made to complete the insertion and the distal aspect of the screw broke at the smooth junction.The distal aspect of the screw was left in the patient but the proximal portion was retrieved.The second pip fusion was completed with a soft tissue reattachment only and the third pip fusion was completed with a k-wire.It was noted that the bone was prepared using ar-4157ds kit and the driver from the ar-4157ds was used for the implant screw.Patient was a (b)(6) male.
|