| Model Number |
SM8 |
| Medical Device Problem Codes |
Decrease in Pressure (1490); Pressure Problem (3012)
|
| Health Effect - Clinical Code |
Hydrocephalus (3272)
|
| Date of Event |
09/06/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Suspision of chronic hypodrainage, despite the valve being in the low position.The valve had to be explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The marks observed on the inlet connector lead us to suspect the use of metallic pliers during the manipulation of the valve.The use of such pliers is inadvisable because it can cause a loosening of the connector, leading to the uncalibration of the valve.The instruction for use of the sm8 valve contains a precaution against the use of metallic pliers.
|
| |
|
Event or Problem Description
|
|
Suspicion of chronic hypodrainage, despite the valve being in the low position.The valve had to be explanted.
|
| |
|