| Model Number |
11.1510 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
09/07/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The dhr review shows the device performed as intended when it was manufactured.During evaluation the device and device remote were found in poor condition.The device and device remote control were able to power on.The device treatment was able to function to specification, frequencies were to specification.The device labeling was reviewed and found to be suitable and adequate for the device to perform its intended use.The device performed to specification.
|
| |
|
Event or Problem Description
|
|
Customer has spinal stenosis and has been wearing the device for two weeks.There were no problems until (b)(6) 2016.She was wearing the unit all day and it had turned itself off.Customer said she "got in the car to go somewhere and all of a sudden it let out a charge that tightened every muscle in my body and caused me to yell out loud." she had her remote control around her neck and was able to decrease the intensity.
|
| |
|