| Catalog Number |
9735015 |
| Medical Device Problem Codes |
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/16/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No parts have been received by the manufacturer for evaluation, thus no analysis can be done at this time.
|
| |
|
Event or Problem Description
|
|
A site representative reported that during an ear, nose, and throat procedure, a malleable suction instrument was unable to be tracked by the navigation system.The instrument tracker was replaced and the issue was resolved.The procedure was completed successfully with the suction and the navigation system after a delay of less than one hour.The patient was not impacted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device expiration date now provided.Device manufacturing date now provided.
|
| |
|