| Model Number |
CMP680300 |
| Medical Device Problem Codes |
Device Emits Odor (1425); Smoking (1585)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/20/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).The field service specialist (fss) visited customer site to inspect the unit.Fss confirmed the visible smoke issue with the unit and replaced the single board computer (sbc) printed circuit board (pcb) to rectify the problem.Fss performed the field service checklist, which the unit passed all functional tests and it was found to meet amo specifications.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
The account reported smell from the system, that visible smoke was observed with the sovereign compact console.There was no patient involvement reported and no patient treatments delayed or cancelled.
|
| |
|
Additional Manufacturer Narrative
|
|
The labeling, trending and risk documentation reviews for this equipment were performed.The trend review shows that there is not a recognizable adverse trend.The risks and mitigations associated with the received complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.A labeling review was conducted; the operator manual for the system was reviewed and found to include adequate instructions for use, warnings and operational errors.All pertinent information available to abbott medical optics has been submitted.
|
| |
|