• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/21/2016
Event Type  malfunction  
Manufacturer Narrative
There was no patient affected.Sorin group (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the s5 gas blender system displayed an error message during a procedure.There was no report of patient injury.The complained gas blender was returned to sorin group (b)(4) for investigation.Visual inspection did not identify any defects or abnormalities.However the reported issue was confirmed during functional testing.The mass-flow meter/controller for air and o2 were replaced and a functional check and new calibration were performed.Functional control and technical safety inspection were carried out and no further issues were discovered.The device was cleaned and disinfected and returned to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related th this type of issue.
 
Event Description
Sorin group (b)(4) received a report that the s5 gas blender system displayed an error message during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6036882
MDR Text Key57733100
Report Number9611109-2016-00673
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 09/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-