AV-TEMECULA-CT ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM; BIORESORBABLE DRUG ELUTING SCAFFOLD Back to Search Results
Catalog Number 1235300-28
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 09/28/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).There was no reported device malfunction and the product was not returned as the scaffold remains in the anatomy.The reported patient effects of dissection, as listed in the absorb gt1 instructions for use is a known adverse event associated with the use of a coronary scaffold in native coronary arteries.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined and the treatment appears to be related to the circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion located in the mid-left anterior descending coronary artery that was moderately calcified.An absorb gt1 2.5 x 18 scaffold was delivered distally with no incidence.Then, an absorb gt1 3.0 x 28 mm scaffold was deployed proximally to the first scaffold when a dissection occurred.A graftmaster 2.8 x 16 mm coronary stent was used as treatment to cover the dissection.There was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSORB GT1 BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Common Device Name
BIORESORBABLE DRUG ELUTING SCAFFOLD
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6038085
Report Number2024168-2016-07041
Device Sequence Number13807941
Product Code PNY
UDI-Device Identifier08717648213069
UDI-Public(01)08717648213069(17)170508(10)6050961
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
P150023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/08/2017
Device Catalogue Number1235300-28
Device Lot Number6050961
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2016
Initial Report FDA Received Date10/18/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OTHER: ABSORB GT1 2.5 X 18 MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-