Catalog Number 6605632
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 09/29/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A field service engineer (fse) evaluated the analyzer on (b)(6) 2016, and the normally closed tubing at valve vl8, the red blood cell (rbc) diluent dispense valve was replaced, resolving the uncontained leak.The beckman coulter internal identifier for this event is (b)(4).
 
Event or Problem Description
A customer reported an uncontained clenz leak of less than 50 ml from a coulter lh 750 hematology analyzer.The customer was wearing personal protective equipment (ppe) consisting of gloves and a laboratory coat at the time the leak was discovered, and there was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
MDR Report Key6040140
Report Number1061932-2016-00878
Device Sequence Number10927766
Product Code GKZ
UDI-Device Identifier15099590264772
UDI-Public(01)15099590264772(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K061574
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number6605632
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2016
Initial Report FDA Received Date10/19/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-