• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

This report is being submitted as part of the remediation for capa (b)(4).

 
Event Description

The customer reported that during the installation of the v60 ventilator, the ventilator failed the ground resistance portion of the electrical safety test. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameV60 VENTILATOR
Type of DeviceVENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key6043963
MDR Text Key58027423
Report Number2031642-2016-02675
Device Sequence Number1
Product Code MNT
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation
Type of Report Initial
Report Date 10/03/2014
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/20/2016
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberV60
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer11/04/2014
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/03/2014
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured09/18/2014
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-