• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCERT MEDICAL, LLC GALEO PRO HF CORONARY GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCERT MEDICAL, LLC GALEO PRO HF CORONARY GUIDEWIRE Back to Search Results
Model Number 512040-001
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/05/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Large sections of ptfe coating had peeled away from the proximal portion of the device; ptfe coating on distal section of device was still properly adhered.Coating adhesion testing performed with no saline soak confirmed that the ptfe adhesion failed at the proximal section of the guidewire and passed at the distal section.Coating adhesion testing performed after 1 hour saline soak found lack of ptfe adhesion at the proximal and distal end of the device.Review of manufacturing and quality records showed that the manufacturing lot had passed the in-process coating adhesion test.There were no other nonconformities observed for this manufacturing lot.Additional testing will be performed on other samples from the same coating lot if they can be obtained.
 
Event or Problem Description
During a complex pci just after dilation of the ostium of the lad it was noticed that the ptfe coating of the guidewire had come off of the part of the wire just outside the guiding catheter.While removing the balloon catheter further parts of the coating came loose.No patient injury or additional intervention was required, and the user did not report that any pieces of the coating were left in the patient.The patient was discharged home following completion of the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALEO PRO HF CORONARY GUIDEWIRE
Common Device Name
CORONARY GUIDEWIRE
Manufacturer (Section D)
CONCERT MEDICAL, LLC
77 accord park drive
norwell MA 02061
MDR Report Key6044059
Report Number1223414-2016-00004
Device Sequence Number902905
Product Code DQX
UDI-Device Identifier00814384020192
UDI-Public00814384020192
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K090193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date05/28/2019
Device Model Number512040-001
Device Catalogue Number389781
Device Lot Number00574827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2016
Initial Report FDA Received Date10/20/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6 FR GUIDING CATHETER; 6 FR TERUMO INTRODUCER SHEATH; BALLOON CATHETER
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
-
-